Ferromagnetic contamination of Ultra-Low-Field-NMR sample containers. Quantification of the problem and possible solutions

30 Jun 2020  ·  Bevilacqua Giuseppe, Biancalana Valerio, Consumi Marco, Dancheva Yordanka, Rossi Claudio, Stiaccini Leonardo, Vigilante Antonio ·

The presence of a weak remanence in Ultra-Low-Field (ULF) NMR sample containers is investigated on the basis of proton precession. The high-sensitivity magnetometer used for the NMR detection, enables simultaneously the measurement of the static field produced in the sample proximity by ferromagnetic contaminants. The presence of the latter is studied by high resolution chemical analyses of the surface, based on X-ray fluorescence spectroscopy and secondary ions mass spectroscopy. Methodologies to reduce the contamination are explored and characterized. This study is of relevance in any ULF-NMR experiment, as in the ULF regime spurious ferromagnetism becomes easily a dominant cause of artefacts.

PDF Abstract
No code implementations yet. Submit your code now

Categories


Instrumentation and Detectors Chemical Physics