Off-axis QEPAS using a pulsed nanosecond Mid-Infrared Optical Parametric Oscillator

11 Aug 2016  ·  Lassen Mikael, Lamard Laurent, Feng Yuyang, peremans Andre, Petersen Jan C. ·

A trace gas sensor, based on quartz-enhanced photoacoustic spectroscopy (QEPAS), consisting of two acoustically coupled micro-resonators (mR) with an off-axis 20 kHz quartz tuning fork (QTF) is demonstrated. The complete acoustically coupled mR system is optimized based on finite element simulations and experimentally verified. The QEPAS sensor is pumped resonantly by a nanosecond pulsed single-mode mid-infrared optical parametric oscillator (MIR OPO). The sensor is used for spectroscopic measurements on methane in the 3.1 um to 3.5 um wavelength region with a resolution bandwidth of 1 cm^-1 and a detection limit of 0.8 ppm. An Allan deviation analysis shows that the detection limit at optimum integration time for the QEPAS sensor is 32 ppbv@190s and that the background noise is solely due to the thermal noise of the QTF.

PDF Abstract
No code implementations yet. Submit your code now

Categories


Instrumentation and Detectors Optics