The Laser calibration of the Atlas Tile Calorimeter during the LHC run 1

12 Oct 2016  ·  system ATLAS Tile Calorimeter ·

This article describes the Laser calibration system of the Atlas hadronic Tile Calorimeter that has been used during the run 1 of the LHC. First, the stability of the system associated readout electronics is studied. It is found to be stable with variations smaller than 0.6 %. Then, the method developed to compute the calibration constants, to correct for the variations of the gain of the calorimeter photomultipliers, is described. These constants were determined with a statistical uncertainty of 0.3 % and a systematic uncertainty of 0.2 % for the central part of the calorimeter and 0.5 % for the end-caps. Finally, the detection and correction of timing mis-configuration of the Tile Calorimeter using the Laser system are also presented.

PDF Abstract
No code implementations yet. Submit your code now

Categories


Instrumentation and Detectors High Energy Physics - Experiment